Filter characteristics
Notch filter for 50-Hz-rejection
-3 dB points at 41 and 59 Hz
Low-pass filter (EEG)
-3 dB point at 49 Hz
Band-pass filter (EMG)
-3 dB points at 16 and 54 Hz
The above two are also anti-aliasing filters for the sampling rate reduction from 512 Hz to 128 Hz.
The following filters are unconditional, i.e., they can not be switched off:
Amplifier time-constant:
1 second, which is a high-pass filter with a
-3 dB point at 0.16 Hz
Hardware anti-aliasing filter
Low-pass with -3 dB point at 102 Hz (in the 512-samples-per-second domain)